DoseRateMode
The dose rate mode of this treatment unit. Supported modes are "HDR", "PDR", "MDR", and "LDR".
Declaration
public string DoseRateMode { get; }
Property Value
Type | Description |
---|---|
System.String |
DwellTimeResolution
The dwell time resolution supported by this treatment unit in seconds.
Declaration
public double DwellTimeResolution { get; }
Property Value
Type | Description |
---|---|
System.Double |
MachineInterface
The interface type for communicating with this brachytherapy treatment unit. Possible types are "GammaMed12i", "GammaMedPlus", "VariSource", "Other", and "Omnitron" (obsolete type).
Declaration
public string MachineInterface { get; }
Property Value
Type | Description |
---|---|
System.String |
MachineModel
The model identifier for this treatment unit. Possible models are "VariSource_5", "VariSource_10", "Remote_Afterloading", "Manual_Loading", "GammaMed12i", and "GammaMedPlus".
Declaration
public string MachineModel { get; }
Property Value
Type | Description |
---|---|
System.String |
MaxDwellTimePerChannel
The maximum combined dwell time in a single channel in seconds.
Declaration
public double MaxDwellTimePerChannel { get; }
Property Value
Type | Description |
---|---|
System.Double |
MaxDwellTimePerPos
The maximum dwell time in a single dwell position in seconds.
Declaration
public double MaxDwellTimePerPos { get; }
Property Value
Type | Description |
---|---|
System.Double |
MaxDwellTimePerTreatment
The maximum combined dwell time in all the channels during a single treatment session. The value is in seconds.
Declaration
public double MaxDwellTimePerTreatment { get; }
Property Value
Type | Description |
---|---|
System.Double |
MaximumChannelLength
The maximum channel length supported by this treatment unit in millimeters.
Declaration
public double MaximumChannelLength { get; }
Property Value
Type | Description |
---|---|
System.Double |
MaximumDwellPositionsPerChannel
The maximum number of dwell positions per channel supported by this treatment unit.
Declaration
public int MaximumDwellPositionsPerChannel { get; }
Property Value
Type | Description |
---|---|
System.Int32 |
MaximumStepSize
The maximum distance between adjacent source positions in millimeters.
Declaration
public double MaximumStepSize { get; }
Property Value
Type | Description |
---|---|
System.Double |
MinAllowedSourcePos
The minimum allowed distance (in millimeters) from the tip of the inner lumen of the applicator to the center of the first dwell position. In other words, no source positions should be placed within this distance from the tip of the inner lumen.
Declaration
public double MinAllowedSourcePos { get; }
Property Value
Type | Description |
---|---|
System.Double |
MinimumChannelLength
The minimum channel length supported by this treatment unit in millimeters.
Declaration
public double MinimumChannelLength { get; }
Property Value
Type | Description |
---|---|
System.Double |
MinimumStepSize
The minimum distance between adjacent source positions in millimeters.
Declaration
public double MinimumStepSize { get; }
Property Value
Type | Description |
---|---|
System.Double |
NumberOfChannels
The number of channels in this treatment unit.
Declaration
public int NumberOfChannels { get; }
Property Value
Type | Description |
---|---|
System.Int32 |
SourceCenterOffsetFromTip
The offset distance (in millimeters) from the tip of the applicator to the center of the source at its first possible dwell position. In other words, the offset accounts for half of the active source length and encapsulation.
Declaration
public double SourceCenterOffsetFromTip { get; }
Property Value
Type | Description |
---|---|
System.Double |
SourceMovementType
The source movement type as defined in DICOM. Possible types are "STEPWISE", "FIXED", and "OSCILLATING".
Declaration
public string SourceMovementType { get; }
Property Value
Type | Description |
---|---|
System.String |
StepSizeResolution
The default step size resolution for this treatment unit in millimeters.
Declaration
public double StepSizeResolution { get; }
Property Value
Type | Description |
---|---|
System.Double |
Methods
GetActiveRadioactiveSource()
Returns the active radioactive source of this treatment unit.
Declaration
public RadioactiveSource GetActiveRadioactiveSource()
Returns
Type | Description |
---|---|
RadioactiveSource | A RadioactiveSource object if the treatment unit has a source installed. Otherwise null. |
WriteXml(XmlWriter)
Converts an object into its XML representation. For more information on how to serialize an object into an XML document, see System.Xml.Serialization.XmlSerializer.
Declaration
public override void WriteXml(XmlWriter writer)