    Represents an external beam plan. For more information, see the definition of the DICOM RT Plan.

    Inheritance
    System.Object
    SerializableObject
    ApiDataObject
    PlanningItem
    PlanSetup
    ExternalPlanSetup
    Implements
    System.Xml.Serialization.IXmlSerializable
    IDVHEstimationCalculator
    Inherited Members
    PlanSetup.GetProtocolPrescriptionsAndMeasures(List<ProtocolPhasePrescription>, List<ProtocolPhaseMeasure>)
    PlanSetup.AddReferencePoint(Boolean, Nullable<VVector>, String)
    PlanSetup.IsValidForPlanApproval(List<PlanValidationResultEsapiDetail>)
    PlanSetup.AddPlanUncertaintyWithParameters(PlanUncertaintyType, Boolean, Double, VVector)
    PlanSetup.ClearCalculationModel(CalculationType)
    PlanSetup.GetCalculationModel(CalculationType)
    PlanSetup.GetCalculationOption(String, String, String)
    PlanSetup.GetCalculationOptions(String)
    PlanSetup.GetDvhEstimationModelName()
    PlanSetup.SetCalculationModel(CalculationType, String)
    PlanSetup.SetCalculationOption(String, String, String)
    PlanSetup.SetPrescription(Int32, DoseValue, Double)
    PlanSetup.SetTargetStructureIfNoDose(Structure, StringBuilder)
    PlanSetup.Id
    PlanSetup.Name
    PlanSetup.Comment
    PlanSetup.PlanNormalizationValue
    PlanSetup.PlanUncertainties
    PlanSetup.PlanObjectiveStructures
    PlanSetup.ApprovalHistory
    PlanSetup.DosePerFractionInPrimaryRefPoint
    PlanSetup.PrescribedDosePerFraction
    PlanSetup.PrescribedPercentage
    PlanSetup.TotalPrescribedDose
    PlanSetup.ApplicationScriptLogs
    PlanSetup.ApprovalStatus
    PlanSetup.BaseDosePlanningItem
    PlanSetup.Beams
    PlanSetup.CreationUserName
    PlanSetup.DosePerFraction
    PlanSetup.DVHEstimates
    PlanSetup.ElectronCalculationModel
    PlanSetup.ElectronCalculationOptions
    PlanSetup.IntegrityHash
    PlanSetup.IsDoseValid
    PlanSetup.IsTreated
    PlanSetup.NumberOfFractions
    PlanSetup.OptimizationSetup
    PlanSetup.PatientSupportDevice
    PlanSetup.PhotonCalculationModel
    PlanSetup.PhotonCalculationOptions
    PlanSetup.PlanIntent
    PlanSetup.PlannedDosePerFraction
    PlanSetup.PlanningApprovalDate
    PlanSetup.PlanningApprover
    PlanSetup.PlanningApproverDisplayName
    PlanSetup.PlanNormalizationMethod
    PlanSetup.PlanNormalizationPoint
    PlanSetup.PlanType
    PlanSetup.PredecessorPlan
    PlanSetup.PredecessorPlanUID
    PlanSetup.PrimaryReferencePoint
    PlanSetup.ProtocolID
    PlanSetup.ProtocolPhaseID
    PlanSetup.ProtonCalculationModel
    PlanSetup.ProtonCalculationOptions
    PlanSetup.ReferencePoints
    PlanSetup.RTPrescription
    PlanSetup.Series
    PlanSetup.SeriesUID
    PlanSetup.TargetVolumeID
    PlanSetup.TotalDose
    PlanSetup.TreatmentApprovalDate
    PlanSetup.TreatmentApprover
    PlanSetup.TreatmentApproverDisplayName
    PlanSetup.TreatmentOrientation
    PlanSetup.TreatmentPercentage
    PlanSetup.TreatmentSessions
    PlanSetup.UID
    PlanSetup.UseGating
    PlanSetup.VerifiedPlan
    PlanningItem.GetClinicalGoals()
    PlanningItem.GetDVHCumulativeData(Structure, DoseValuePresentation, VolumePresentation, Double)
    PlanningItem.GetDoseAtVolume(Structure, Double, VolumePresentation, DoseValuePresentation)
    PlanningItem.GetVolumeAtDose(Structure, DoseValue, VolumePresentation)
    PlanningItem.Course
    PlanningItem.CreationDateTime
    PlanningItem.Dose
    PlanningItem.DoseValuePresentation
    PlanningItem.StructureSet
    PlanningItem.StructuresSelectedForDvh
    ApiDataObject.ToString()
    ApiDataObject.Equals(Object)
    ApiDataObject.GetHashCode()
    ApiDataObject.HistoryUserName
    ApiDataObject.HistoryUserDisplayName
    ApiDataObject.HistoryDateTime
    SerializableObject.GetSchema()
    SerializableObject.ReadXml(XmlReader)
    SerializableObject.ClearSerializationHistory()
    System.Object.Equals(System.Object, System.Object)
    System.Object.ReferenceEquals(System.Object, System.Object)
    System.Object.GetType()
    System.Object.MemberwiseClone()
    Namespace: VMS.TPS.Common.Model.API
    Assembly: VMS.TPS.Common.Model.API.dll
    Syntax
    public class ExternalPlanSetup : PlanSetup, IXmlSerializable, IDVHEstimationCalculator

    Properties

    DoseAsEvaluationDose

    The evaluation dose is connected to the plan and contains voxels that are set by SetVoxels(Int32, Int32[,]) instead of the dose calculation algorithm. Returns null if the plan has no evaluation dose.

    Declaration
    public EvaluationDose DoseAsEvaluationDose { get; }
    Property Value
    Type Description
    EvaluationDose

    TradeoffExplorationContext

    [Availability of this property depends on your Eclipse Scripting API license] Gets the TradeoffExplorationContext.

    Declaration
    public TradeoffExplorationContext TradeoffExplorationContext { get; }
    Property Value
    Type Description
    TradeoffExplorationContext
    Remarks

    To be able to get the TradeoffExplorationContext class, the plan needs to have the optimization objectives set.

    Methods

    AddArcBeam(ExternalBeamMachineParameters, VRect<Double>, Double, Double, Double, GantryDirection, Double, VVector)

    [Availability of this method depends on your Eclipse Scripting API license] Adds a new arc beam to the plan.

    Declaration
    public Beam AddArcBeam(ExternalBeamMachineParameters machineParameters, VRect<double> jawPositions, double collimatorAngle, double gantryAngle, double gantryStop, GantryDirection gantryDirection, double patientSupportAngle, VVector isocenter)
    Parameters
    Type Name Description
    ExternalBeamMachineParameters machineParameters

    Configuration of the used treatment unit. See ExternalBeamMachineParameters.

    VRect<System.Double> jawPositions

    The jaw positions of the treatment beam in millimeters, and in IEC BEAM LIMITING DEVICE coordinates.

    System.Double collimatorAngle

    The orientation of the IEC BEAM LIMITING DEVICE coordinate system with respect to the IEC GANTRY coordinate system (in degrees).

    System.Double gantryAngle

    The orientation of the IEC GANTRY coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    System.Double gantryStop

    The stop angle of the gantry, in degrees.

    GantryDirection gantryDirection

    The rotation direction of the arc.

    System.Double patientSupportAngle

    The orientation of the IEC PATIENT SUPPORT (turntable) coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    VVector isocenter

    The isocenter position, in millimeters.

    Returns
    Type Description
    Beam

    The created Beam.

    Remarks

    All angles must be specified in the range [0, 360[.

    AddConformalArcBeam(ExternalBeamMachineParameters, Double, Int32, Double, Double, GantryDirection, Double, VVector)

    [Availability of this method depends on your Eclipse Scripting API license] Adds a new conformal arc beam to the plan.

    Declaration
    public Beam AddConformalArcBeam(ExternalBeamMachineParameters machineParameters, double collimatorAngle, int controlPointCount, double gantryAngle, double gantryStop, GantryDirection gantryDirection, double patientSupportAngle, VVector isocenter)
    Parameters
    Type Name Description
    ExternalBeamMachineParameters machineParameters

    Configuration of the used treatment unit. See ExternalBeamMachineParameters.

    System.Double collimatorAngle

    The orientation of the IEC BEAM LIMITING DEVICE coordinate system with respect to the IEC GANTRY coordinate system (in degrees).

    System.Int32 controlPointCount

    The number of control points to use.

    System.Double gantryAngle

    The orientation of the IEC GANTRY coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    System.Double gantryStop

    The stop angle of the gantry, in degrees.

    GantryDirection gantryDirection

    The rotation direction of the arc.

    System.Double patientSupportAngle

    The orientation of the IEC PATIENT SUPPORT (turntable) coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    VVector isocenter

    The isocenter position, in millimeters.

    Returns
    Type Description
    Beam

    The created beam.

    Remarks

    All angles must be specified in the range [0, 360[.

    AddFixedSequenceBeam(ExternalBeamMachineParameters, Double, Double, VVector)

    [Availability of this method depends on your Eclipse Scripting API license] Adds a new fixed sequence (flattening sequence) beam to the plan.

    Declaration
    public Beam AddFixedSequenceBeam(ExternalBeamMachineParameters machineParameters, double collimatorAngle, double gantryAngle, VVector isocenter)
    Parameters
    Type Name Description
    ExternalBeamMachineParameters machineParameters

    Configuration of the used treatment unit. See ExternalBeamMachineParameters. The machine must be Halcyon.

    System.Double collimatorAngle

    The orientation of the IEC BEAM LIMITING DEVICE coordinate system with respect to the IEC GANTRY coordinate system (in degrees).

    System.Double gantryAngle

    The orientation of the IEC GANTRY coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    VVector isocenter

    The isocenter position, in millimeters.

    Returns
    Type Description
    Beam

    The created beam.

    AddImagingSetup(ExternalBeamMachineParameters, ImagingBeamSetupParameters, Structure)

    [Availability of this method depends on your Eclipse Scripting API license] Adds new imaging beam or beams to the plan. First removes the existing imaging beams. If MV imaging beams are involved, the plan should be optimized again.

    Declaration
    public bool AddImagingSetup(ExternalBeamMachineParameters machineParameters, ImagingBeamSetupParameters setupParameters, Structure targetStructure)
    Parameters
    Type Name Description
    ExternalBeamMachineParameters machineParameters

    Configuration of the used treatment unit. See ExternalBeamMachineParameters.

    ImagingBeamSetupParameters setupParameters

    Setup parameters for the imaging beam. See ImagingBeamSetupParameters.

    Structure targetStructure

    Target structure for imaging. If not specified, the target structure in the plan is used.

    Returns
    Type Description
    System.Boolean

    True if imaging beams were successfully created.

    AddMLCArcBeam(ExternalBeamMachineParameters, Single[,], VRect<Double>, Double, Double, Double, GantryDirection, Double, VVector)

    [Availability of this method depends on your Eclipse Scripting API license] Adds a new arc beam with an MLC to the plan.

    Declaration
    public Beam AddMLCArcBeam(ExternalBeamMachineParameters machineParameters, float[, ] leafPositions, VRect<double> jawPositions, double collimatorAngle, double gantryAngle, double gantryStop, GantryDirection gantryDirection, double patientSupportAngle, VVector isocenter)
    Parameters
    Type Name Description
    ExternalBeamMachineParameters machineParameters

    Configuration of the used treatment unit. See ExternalBeamMachineParameters.

    System.Single[,] leafPositions

    The positions of the beam collimator leaf pairs in millimeters, and in the IEC BEAMLIMITING DEVICE coordinate axis appropriate to the MLC device type. For example, the X-axis for MLCX and the Y-axis for MLCY. The two-dimensional array is indexed [bank, leaf], where the bank is either 0 or 1. Bank 0 represents the leaf bank to the negative MLC X direction, and bank 1 to the positive MLC X direction. Passing null selects the default.

    VRect<System.Double> jawPositions

    The jaw positions of the treatment beam in millimeters, and in IEC BEAM LIMITING DEVICE coordinates.

    System.Double collimatorAngle

    The orientation of the IEC BEAM LIMITING DEVICE coordinate system with respect to the IEC GANTRY coordinate system (in degrees).

    System.Double gantryAngle

    The orientation of the IEC GANTRY coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    System.Double gantryStop

    The stop angle of the gantry, in degrees.

    GantryDirection gantryDirection

    The direction of the arc.

    System.Double patientSupportAngle

    The orientation of the IEC PATIENT SUPPORT (turntable) coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    VVector isocenter

    The isocenter position, in millimeters.

    Returns
    Type Description
    Beam

    The created beam.

    Remarks

    All angles must be specified in the range [0, 360[.

    AddMLCBeam(ExternalBeamMachineParameters, Single[,], VRect<Double>, Double, Double, Double, VVector)

    [Availability of this method depends on your Eclipse Scripting API license] Adds a new beam with an MLC to the plan.

    Declaration
    public Beam AddMLCBeam(ExternalBeamMachineParameters machineParameters, float[, ] leafPositions, VRect<double> jawPositions, double collimatorAngle, double gantryAngle, double patientSupportAngle, VVector isocenter)
    Parameters
    Type Name Description
    ExternalBeamMachineParameters machineParameters

    Configuration of the used treatment unit. See ExternalBeamMachineParameters.

    System.Single[,] leafPositions

    The positions of the beam collimator leaf pairs in millimeters, and in the IEC BEAMLIMITING DEVICE coordinate axis appropriate to the MLC device type. For example, the X-axis for MLCX and the Y-axis for MLCY. The two-dimensional array is indexed [bank, leaf] where the bank is either 0 or 1. Bank 0 represents the leaf bank to the negative MLC X direction, and bank 1 to the positive MLC X direction. Passing null selects the default.

    VRect<System.Double> jawPositions

    The jaw positions of the treatment beam in millimeters, and in IEC BEAM LIMITING DEVICE coordinates.

    System.Double collimatorAngle

    The orientation of the IEC BEAM LIMITING DEVICE coordinate system with respect to the IEC GANTRY coordinate system (in degrees).

    System.Double gantryAngle

    The orientation of the IEC GANTRY coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    System.Double patientSupportAngle

    The orientation of the IEC PATIENT SUPPORT (turntable) coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    VVector isocenter

    The isocenter position, in millimeters.

    Returns
    Type Description
    Beam

    The created beam.

    Remarks

    All angles must be specified in the range [0, 360[.

    AddMLCSetupBeam(ExternalBeamMachineParameters, Single[,], VRect<Double>, Double, Double, Double, VVector)

    [Availability of this method depends on your Eclipse Scripting API license] Adds a new static setup field with an MLC to the plan.

    Declaration
    public Beam AddMLCSetupBeam(ExternalBeamMachineParameters machineParameters, float[, ] leafPositions, VRect<double> jawPositions, double collimatorAngle, double gantryAngle, double patientSupportAngle, VVector isocenter)
    Parameters
    Type Name Description
    ExternalBeamMachineParameters machineParameters

    Configuration of the used treatment unit. See ExternalBeamMachineParameters.

    System.Single[,] leafPositions

    The positions of the beam collimator leaf pairs in millimeters, and in the IEC BEAMLIMITING DEVICE coordinate axis appropriate to the MLC device type. For example, the X-axis for MLCX and the Y-axis for MLCY. The two-dimensional array is indexed [bank, leaf] where the bank is either 0 or 1. Bank 0 represents the leaf bank to the negative MLC X direction, and bank 1 to the positive MLC X direction. Passing null selects the default.

    VRect<System.Double> jawPositions

    The jaw positions of the setup field in millimeters, and in IEC BEAM LIMITING DEVICE coordinates.

    System.Double collimatorAngle

    The orientation of the IEC BEAM LIMITING DEVICE coordinate system with respect to the IEC GANTRY coordinate system (in degrees).

    System.Double gantryAngle

    The orientation of the IEC GANTRY coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    System.Double patientSupportAngle

    The orientation of the IEC PATIENT SUPPORT (turntable) coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    VVector isocenter

    The isocenter position, in millimeters.

    Returns
    Type Description
    Beam

    The created beam.

    Remarks

    All angles must be specified in the range [0, 360[.

    AddMultipleStaticSegmentBeam(ExternalBeamMachineParameters, IEnumerable<Double>, Double, Double, Double, VVector)

    [Availability of this method depends on your Eclipse Scripting API license] Adds a new IMRT beam to the plan and defines Multiple Static Segments as the delivery technique.

    Declaration
    public Beam AddMultipleStaticSegmentBeam(ExternalBeamMachineParameters machineParameters, IEnumerable<double> metersetWeights, double collimatorAngle, double gantryAngle, double patientSupportAngle, VVector isocenter)
    Parameters
    Type Name Description
    ExternalBeamMachineParameters machineParameters

    Configuration of the used treatment unit. See ExternalBeamMachineParameters.

    System.Collections.Generic.IEnumerable<System.Double> metersetWeights

    The meterset weight for each control point of the beam. The number of meterset weight items defines the number of created control points.

    System.Double collimatorAngle

    The orientation of the IEC BEAM LIMITING DEVICE coordinate system with respect to the IEC GANTRY coordinate system (in degrees).

    System.Double gantryAngle

    The orientation of the IEC GANTRY coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    System.Double patientSupportAngle

    The orientation of the IEC PATIENT SUPPORT (turntable) coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    VVector isocenter

    The isocenter position, in millimeters.

    Returns
    Type Description
    Beam

    The created beam.

    Remarks

    All angles must be specified in the range [0, 360[.

    AddSetupBeam(ExternalBeamMachineParameters, VRect<Double>, Double, Double, Double, VVector)

    [Availability of this method depends on your Eclipse Scripting API license] Adds a new static setup field to the plan.

    Declaration
    public Beam AddSetupBeam(ExternalBeamMachineParameters machineParameters, VRect<double> jawPositions, double collimatorAngle, double gantryAngle, double patientSupportAngle, VVector isocenter)
    Parameters
    Type Name Description
    ExternalBeamMachineParameters machineParameters

    Configuration of the used treatment unit. See ExternalBeamMachineParameters.

    VRect<System.Double> jawPositions

    The jaw positions of the setup field in millimeters, and in IEC BEAM LIMITING DEVICE coordinates.

    System.Double collimatorAngle

    The orientation of the IEC BEAM LIMITING DEVICE coordinate system with respect to the IEC GANTRY coordinate system (in degrees).

    System.Double gantryAngle

    The orientation of the IEC GANTRY coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    System.Double patientSupportAngle

    The orientation of the IEC PATIENT SUPPORT (turntable) coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    VVector isocenter

    The isocenter position, in millimeters.

    Returns
    Type Description
    Beam

    The created beam.

    Remarks

    All angles must be specified in the range [0, 360[.

    AddSlidingWindowBeam(ExternalBeamMachineParameters, IEnumerable<Double>, Double, Double, Double, VVector)

    [Availability of this method depends on your Eclipse Scripting API license] Adds a new IMRT beam to the plan and defines Sliding Window as the delivery technique.

    Declaration
    public Beam AddSlidingWindowBeam(ExternalBeamMachineParameters machineParameters, IEnumerable<double> metersetWeights, double collimatorAngle, double gantryAngle, double patientSupportAngle, VVector isocenter)
    Parameters
    Type Name Description
    ExternalBeamMachineParameters machineParameters

    Configuration of the used treatment unit. See ExternalBeamMachineParameters.

    System.Collections.Generic.IEnumerable<System.Double> metersetWeights

    The meterset weight for each control point of the beam. The number of meterset weight items defines the number of created control points.

    System.Double collimatorAngle

    The orientation of the IEC BEAM LIMITING DEVICE coordinate system with respect to the IEC GANTRY coordinate system (in degrees).

    System.Double gantryAngle

    The orientation of the IEC GANTRY coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    System.Double patientSupportAngle

    The orientation of the IEC PATIENT SUPPORT (turntable) coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    VVector isocenter

    The isocenter position, in millimeters.

    Returns
    Type Description
    Beam

    The created beam.

    Remarks

    All angles must be specified in the range [0, 360[.

    AddSlidingWindowBeamForFixedJaws(ExternalBeamMachineParameters, IEnumerable<Double>, Double, Double, Double, VVector)

    [Availability of this method depends on your Eclipse Scripting API license] Adds a new sliding window IMRT beam and takes jaw positions from machine configuration. In case of a Halcyon machine, the special leaf position requirements are taken into account.

    Declaration
    public Beam AddSlidingWindowBeamForFixedJaws(ExternalBeamMachineParameters machineParameters, IEnumerable<double> metersetWeights, double collimatorAngle, double gantryAngle, double patientSupportAngle, VVector isocenter)
    Parameters
    Type Name Description
    ExternalBeamMachineParameters machineParameters

    Configuration of the used treatment unit. See ExternalBeamMachineParameters.

    System.Collections.Generic.IEnumerable<System.Double> metersetWeights

    The meterset weight for each control point of the beam. The number of meterset weight items defines the number of created control points.

    System.Double collimatorAngle

    The orientation of the IEC BEAM LIMITING DEVICE coordinate system with respect to the IEC GANTRY coordinate system (in degrees).

    System.Double gantryAngle

    Gantry angle in the IEC GANTRY coordinate system (in degrees).

    System.Double patientSupportAngle

    The orientation of the IEC PATIENT SUPPORT (turntable) coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    VVector isocenter

    The isocenter position in millimeters.

    Returns
    Type Description
    Beam

    The created beam.

    Remarks

    All angles must be specified in the range [0, 360[.

    AddStaticBeam(ExternalBeamMachineParameters, VRect<Double>, Double, Double, Double, VVector)

    [Availability of this method depends on your Eclipse Scripting API license] Adds a new static beam to the plan.

    Declaration
    public Beam AddStaticBeam(ExternalBeamMachineParameters machineParameters, VRect<double> jawPositions, double collimatorAngle, double gantryAngle, double patientSupportAngle, VVector isocenter)
    Parameters
    Type Name Description
    ExternalBeamMachineParameters machineParameters

    Configuration of the used treatment unit. See ExternalBeamMachineParameters.

    VRect<System.Double> jawPositions

    The jaw positions of the treatment beam in millimeters, and in IEC BEAM LIMITING DEVICE coordinates.

    System.Double collimatorAngle

    The orientation of the IEC BEAM LIMITING DEVICE coordinate system with respect to the IEC GANTRY coordinate system (in degrees).

    System.Double gantryAngle

    The orientation of the IEC GANTRY coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    System.Double patientSupportAngle

    The orientation of the IEC PATIENT SUPPORT (turntable) coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    VVector isocenter

    The isocenter position, in millimeters.

    Returns
    Type Description
    Beam

    The created beam.

    Remarks

    All angles must be specified in the range [0, 360[.

    AddVMATBeam(ExternalBeamMachineParameters, IEnumerable<Double>, Double, Double, Double, GantryDirection, Double, VVector)

    [Availability of this method depends on your Eclipse Scripting API license] Adds a new VMAT beam to the plan.

    Declaration
    public Beam AddVMATBeam(ExternalBeamMachineParameters machineParameters, IEnumerable<double> metersetWeights, double collimatorAngle, double gantryAngle, double gantryStop, GantryDirection gantryDirection, double patientSupportAngle, VVector isocenter)
    Parameters
    Type Name Description
    ExternalBeamMachineParameters machineParameters

    Configuration of the used treatment unit. See ExternalBeamMachineParameters.

    System.Collections.Generic.IEnumerable<System.Double> metersetWeights

    The meterset weight for each control point of the beam. The number of meterset weight items defines the number of created control points.

    System.Double collimatorAngle

    The orientation of the IEC BEAM LIMITING DEVICE coordinate system with respect to the IEC GANTRY coordinate system (in degrees).

    System.Double gantryAngle

    The orientation of the IEC GANTRY coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    System.Double gantryStop

    The stop angle of the gantry, in degrees.

    GantryDirection gantryDirection

    The rotation direction of the arc.

    System.Double patientSupportAngle

    The orientation of the IEC PATIENT SUPPORT (turntable) coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    VVector isocenter

    The isocenter position, in millimeters.

    Returns
    Type Description
    Beam

    The created beam.

    Remarks

    All angles must be specified in the range [0, 360[.

    AddVMATBeamForFixedJaws(ExternalBeamMachineParameters, IEnumerable<Double>, Double, Double, Double, GantryDirection, Double, VVector)

    [Availability of this method depends on your Eclipse Scripting API license] Adds a new VMAT beam and takes jaw positions from machine configuration. In case of a Halcyon machine, the special leaf position requirements are taken into account.

    Declaration
    public Beam AddVMATBeamForFixedJaws(ExternalBeamMachineParameters machineParameters, IEnumerable<double> metersetWeights, double collimatorAngle, double gantryStartAngle, double gantryStopAngle, GantryDirection gantryDir, double patientSupportAngle, VVector isocenter)
    Parameters
    Type Name Description
    ExternalBeamMachineParameters machineParameters

    Configuration of the used treatment unit. See ExternalBeamMachineParameters.

    System.Collections.Generic.IEnumerable<System.Double> metersetWeights

    The meterset weight for each control point of the beam. The number of meterset weight items defines the number of created control points.

    System.Double collimatorAngle

    The orientation of the IEC BEAM LIMITING DEVICE coordinate system with respect to the IEC GANTRY coordinate system (in degrees).

    System.Double gantryStartAngle

    Gantry start angle in the IEC GANTRY coordinate system (in degrees).

    System.Double gantryStopAngle

    Gantry stop angle in the IEC GANTRY coordinate system (in degrees).

    GantryDirection gantryDir

    Gantry direction

    System.Double patientSupportAngle

    The orientation of the IEC PATIENT SUPPORT (turntable) coordinate system with respect to the IEC FIXED REFERENCE coordinate system (in degrees).

    VVector isocenter

    The isocenter position in millimeters.

    Returns
    Type Description
    Beam

    The created beam.

    Remarks

    All angles must be specified in the range [0, 360[.

    CalculateDose()

    [Availability of this method depends on your Eclipse Scripting API license] Calculates the dose for the plan.

    Declaration
    public CalculationResult CalculateDose()
    Returns
    Type Description
    CalculationResult

    The calculation result. See calculation details from CalculationLogs.

    CalculateDoseWithPresetValues(List<KeyValuePair<String, MetersetValue>>)

    Calculates the dose for the plan with preset meterset values. This applies only to IMRT beams.

    Declaration
    public CalculationResult CalculateDoseWithPresetValues(List<KeyValuePair<string, MetersetValue>> presetValues)
    Parameters
    Type Name Description
    System.Collections.Generic.List<System.Collections.Generic.KeyValuePair<System.String, MetersetValue>> presetValues

    The list of preset meterset values for the beams. An element of the list should contain a pair consisting of the beam identifier (as the key) and the corresponding preset meterset value for the beam. Each beam in the plan must have one entry in the list.

    Returns
    Type Description
    CalculationResult

    The calculation result. See calculation details from CalculationLogs.

    CalculateDVHEstimates(String, Dictionary<String, DoseValue>, Dictionary<String, String>)

    [Availability of this method depends on your Eclipse Scripting API license] Calculates DVH estimates for the plan.

    Declaration
    public CalculationResult CalculateDVHEstimates(string modelId, Dictionary<string, DoseValue> targetDoseLevels, Dictionary<string, string> structureMatches)
    Parameters
    Type Name Description
    System.String modelId

    DVH estimation model ID.

    System.Collections.Generic.Dictionary<System.String, DoseValue> targetDoseLevels

    Dose levels for the target structures.

    System.Collections.Generic.Dictionary<System.String, System.String> structureMatches

    Matching between the IDs of the structures contained in the structure set used in the plan (the Key in the Dictionary) and the structures used in the DVH estimation model (the Value in the Dictionary).

    Returns
    Type Description
    CalculationResult

    The calculation result. See the calculation details CalculationLogs

    CalculateLeafMotions()

    [Availability of this method depends on your Eclipse Scripting API license] Calculates leaf motions using the calculation options of the plan setup. Before calling this method, set the calculation models for leaf motions and dose calculation.

    Declaration
    public CalculationResult CalculateLeafMotions()
    Returns
    Type Description
    CalculationResult

    The calculation result. See calculation details from CalculationLogs.

    See Also
    SetCalculationModel(CalculationType, System.String)
    SetCalculationOption(System.String, System.String, System.String)

    CalculateLeafMotions(LMCMSSOptions)

    [Availability of this method depends on your Eclipse Scripting API license] Calculates leaf motions using the given calculation options, which override the calculation options defined for the plan setup. Before calling this method, set the calculation models for leaf motions and dose calculation. The used leaf motion calculation algorithm must be MSS Leaf Motion Calculator (LMCMSS).

    Declaration
    public CalculationResult CalculateLeafMotions(LMCMSSOptions options)
    Parameters
    Type Name Description
    LMCMSSOptions options

    Calculation options for the MSS Leaf Motion Calculator (LMCMSS) algorithm.

    Returns
    Type Description
    CalculationResult

    The calculation result. See calculation details from CalculationLogs.

    See Also
    SetCalculationModel(CalculationType, System.String)
    SetCalculationOption(System.String, System.String, System.String)

    CalculateLeafMotions(LMCVOptions)

    [Availability of this method depends on your Eclipse Scripting API license] Calculates leaf motions using the given calculation options, which override the calculation options defined for the plan setup. Before calling this method, set the calculation models for leaf motions and dose calculation. The used leaf motion calculation algorithm must be Varian Leaf Motion Calculator (LMCV).

    Declaration
    public CalculationResult CalculateLeafMotions(LMCVOptions options)
    Parameters
    Type Name Description
    LMCVOptions options

    Calculation options for the Varian Leaf Motion Calculator (LMCV) algorithm.

    Returns
    Type Description
    CalculationResult

    The calculation result. See calculation details from CalculationLogs.

    See Also
    SetCalculationModel(CalculationType, System.String)
    SetCalculationOption(System.String, System.String, System.String)

    CalculateLeafMotions(SmartLMCOptions)

    Calculates leaf motions using the given calculation options, which override the calculation options defined for the plan setup. Before calling this method, set the calculation models for leaf motions and dose calculation. The used leaf motion calculation algorithm must be Varian Smart LMC.

    Declaration
    public CalculationResult CalculateLeafMotions(SmartLMCOptions options)
    Parameters
    Type Name Description
    SmartLMCOptions options

    Calculation options for the Varian Smart LMC algorithm.

    Returns
    Type Description
    CalculationResult

    The calculation result. See calculation details from CalculationLogs.

    See Also
    SetCalculationModel(CalculationType, System.String)
    SetCalculationOption(System.String, System.String, System.String)

    CalculateLeafMotionsAndDose()

    Calculate leaf motions and dose using the calculation models defined in the plan setup.

    Declaration
    public CalculationResult CalculateLeafMotionsAndDose()
    Returns
    Type Description
    CalculationResult

    Result of the intermediate dose calculation.

    CalculatePlanUncertaintyDoses()

    [Availability of this method depends on your Eclipse Scripting API license] Calculates the plan uncertainty dose for the photon plan.

    Declaration
    public CalculationResult CalculatePlanUncertaintyDoses()
    Returns
    Type Description
    CalculationResult

    The calculation result. See calculation details in CalculationLogs.

    CopyEvaluationDose(Dose)

    [Availability of this method depends on your Eclipse Scripting API license] Creates an evaluation dose for the plan. The voxels in an evaluation dose are copied from the existing dose matrix. To create an evaluation dose, the plan must not contain any beams.

    Declaration
    public EvaluationDose CopyEvaluationDose(Dose existing)
    Parameters
    Type Name Description
    Dose existing

    The dose matrix is copied from this object.

    Returns
    Type Description
    EvaluationDose

    A new evaluation dose object.

    CreateEvaluationDose()

    [Availability of this method depends on your Eclipse Scripting API license] Creates an evaluation dose for the plan. The voxels in an evaluation dose can be set using the Eclipse Scripting API instead of a dose calculation algorithm. To create an evaluation dose, the plan must not contain any beams. To set the evaluation dose voxels, retrieve the dose matrix using the DoseAsEvaluationDose property of the plan and then use SetVoxels(Int32, Int32[,]).

    Declaration
    public EvaluationDose CreateEvaluationDose()
    Returns
    Type Description
    EvaluationDose

    A new evaluation dose object.

    Remarks

    Saving modifications to the database is not possible if the evaluation dose has been created but voxels have not been set.

    GetModelsForCalculationType(CalculationType)

    Returns a list of calculation models for the given calculation type.

    Declaration
    public IEnumerable<string> GetModelsForCalculationType(CalculationType calculationType)
    Parameters
    Type Name Description
    CalculationType calculationType

    The type of calculation.

    Returns
    Type Description
    System.Collections.Generic.IEnumerable<System.String>

    The list of calculation models.

    Optimize()

    Runs IMRT optimization for the plan setup. The Multileaf Collimator (MLC) is determined automatically. If there are more than one MLC or no MLC at all, an exception is thrown. Plan normalization method is changed to 'No plan normalization' after successful optimization.

    Declaration
    public OptimizerResult Optimize()
    Returns
    Type Description
    OptimizerResult

    The result of the optimization.

    Optimize(Int32)

    Runs IMRT optimization for the plan setup. All existing optimal fluences are removed from the beams. Fails if the method cannot select an MLC for the beams. Plan normalization method is changed to 'No plan normalization' after successful optimization.

    Declaration
    public OptimizerResult Optimize(int maxIterations)
    Parameters
    Type Name Description
    System.Int32 maxIterations

    The maximum number of iterations.

    Returns
    Type Description
    OptimizerResult

    The result of the optimization.

    Optimize(Int32, OptimizationOption)

    Runs IMRT optimization for the plan setup. Fails if the method cannot select an MLC for the beams. Plan normalization method is changed to 'No plan normalization' after successful optimization.

    Declaration
    public OptimizerResult Optimize(int maxIterations, OptimizationOption optimizationOption)
    Parameters
    Type Name Description
    System.Int32 maxIterations

    The maximum number of iterations.

    OptimizationOption optimizationOption

    Optimization options.

    Returns
    Type Description
    OptimizerResult

    The result of the optimization.

    Optimize(Int32, OptimizationOption, String)

    Runs IMRT optimization for the plan setup. You can add the Multileaf Collimator (MLC) as a parameter. Plan normalization method is changed to 'No plan normalization' after successful optimization.

    Declaration
    public OptimizerResult Optimize(int maxIterations, OptimizationOption optimizationOption, string mlcId)
    Parameters
    Type Name Description
    System.Int32 maxIterations

    The maximum number of iterations.

    OptimizationOption optimizationOption

    Optimization option specifying the initial state.

    System.String mlcId

    The identifier for the Multileaf Collimator (MLC).

    Returns
    Type Description
    OptimizerResult

    The result of the optimization.

    Optimize(OptimizationOptionsIMRT)

    [Availability of this method depends on your Eclipse Scripting API license] Runs IMRT optimization for the plan setup. Plan normalization method is changed to 'No plan normalization' after successful optimization.

    Declaration
    public OptimizerResult Optimize(OptimizationOptionsIMRT options)
    Parameters
    Type Name Description
    OptimizationOptionsIMRT options

    Options specifying initial state, intermediate dose, and termination upon convergence.

    Returns
    Type Description
    OptimizerResult

    Result of the optimization.

    OptimizeVMAT()

    [Availability of this method depends on your Eclipse Scripting API license] Runs VMAT optimization for the plan setup. The Multileaf Collimator (MLC) is determined automatically. If there are more than one MLC or no MLC at all, an exception is thrown. Plan normalization method is changed to 'No plan normalization' after successful optimization.

    Declaration
    public OptimizerResult OptimizeVMAT()
    Returns
    Type Description
    OptimizerResult

    The result of the optimization.

    OptimizeVMAT(String)

    [Availability of this method depends on your Eclipse Scripting API license] Runs VMAT optimization for the plan setup. You can add the MLC as a parameter. Plan normalization method is changed to 'No plan normalization' after successful optimization.

    Declaration
    public OptimizerResult OptimizeVMAT(string mlcId)
    Parameters
    Type Name Description
    System.String mlcId

    The identifier for the Multileaf Collimator (MLC).

    Returns
    Type Description
    OptimizerResult

    The result of the optimization.

    OptimizeVMAT(OptimizationOptionsVMAT)

    [Availability of this method depends on your Eclipse Scripting API license] Runs VMAT optimization for the plan setup using intermediate dose calculation. Plan normalization method is changed to 'No plan normalization' after successful optimization.

    Declaration
    public OptimizerResult OptimizeVMAT(OptimizationOptionsVMAT options)
    Parameters
    Type Name Description
    OptimizationOptionsVMAT options

    VMAT optimization options.

    Returns
    Type Description
    OptimizerResult

    The result of the optimization.

    RemoveBeam(Beam)

    [Availability of this method depends on your Eclipse Scripting API license] Removes a beam from the plan.

    Declaration
    public void RemoveBeam(Beam beam)
    Parameters
    Type Name Description
    Beam beam

    The beam that will be removed from the plan.

    WriteXml(XmlWriter)

    Converts an object into its XML representation. For more information on how to serialize an object into an XML document, see System.Xml.Serialization.XmlSerializer.

    Declaration
    public override void WriteXml(XmlWriter writer)
    Parameters
    Type Name Description
    System.Xml.XmlWriter writer

    The System.Xml.XmlWriter stream, to which the object is serialized.

    Overrides
    PlanSetup.WriteXml(XmlWriter)

    Implements

    System.Xml.Serialization.IXmlSerializable
    IDVHEstimationCalculator
    In This Article